Tr. Mccauley et al., VALIDATION OF VOLUME FLOW MEASUREMENTS WITH CINE PHASE-CONTRAST MR-IMAGING FOR PERIPHERAL ARTERIAL WAVE-FORMS, Journal of magnetic resonance imaging, 5(6), 1995, pp. 663-668
A now phantom was used to study MR volume now measurements for monopha
sic and triphasic waveforms over the now range expected in peripheral
arteries at rest and with exercise (2-24 mL/sec, n = 50). The improvem
ent in accuracy with phase-correction image processing to eliminate er
rors caused by eddy currents was measured, Volume now estimates with D
oppler sonography were also measured, MR volume now measurements corre
lated with volume collection with r = .996 and mean error = 4.6%, Phas
e-correction processing decreased mean error from 12.6% to 4.6% (P < .
001, paired t-test), Doppler sonography had a higher mean error of 10.
3% (P < .001, unpaired t-test). Cine phase-contrast MR imaging provide
s accurate estimates of volume blood now for waveforms and now ranges
expected in peripheral arteries.