THE HIGH MOMENTUM SPECTROMETER DRIFT CHAMBERS IN HALL-C AT CEBAF

Citation
Ok. Baker et al., THE HIGH MOMENTUM SPECTROMETER DRIFT CHAMBERS IN HALL-C AT CEBAF, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 367(1-3), 1995, pp. 92-95
Citations number
11
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
367
Issue
1-3
Year of publication
1995
Pages
92 - 95
Database
ISI
SICI code
0168-9002(1995)367:1-3<92:THMSDC>2.0.ZU;2-6
Abstract
The multiwire drift chambers to be used in the High Momentum Spectrome ter (HMS) at the Continuous Electron Beam Accelerator Facility (CEBAF) have been designed and constructed, and recently employed in initial data-taking runs. These chambers are used to reconstruct scattered cha rged particle momenta in the HMS using C-12 and BeO2 targets for incid ent electron energies up to 2.2 GeV. Offline analysis of the data indi cate that these drift chambers have spatial resolution (per plane) of about 115 mu m (sigma) in rates approaching a kHz/wire/mm. It is expec ted that this performance will improve at higher momenta where multipl e scattering contributions are smaller.