RAPID AND ECONOMICAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF NORFLOXACININ SERUM USING A MICROPARTICULATE C-18 GUARD CARTRIDGE

Citation
Sc. Wallis et al., RAPID AND ECONOMICAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF NORFLOXACININ SERUM USING A MICROPARTICULATE C-18 GUARD CARTRIDGE, Journal of chromatography B. Biomedical applications, 674(2), 1995, pp. 306-309
Citations number
14
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical applications
ISSN journal
15726495 → ACNP
Volume
674
Issue
2
Year of publication
1995
Pages
306 - 309
Database
ISI
SICI code
Abstract
A rapid and economical high-performance liquid chromatographic assay i s described for norfloxacin in serum. Samples (100 mu l) containing N- ethylnorfloxacin as the internal standard were extracted into 1 ml of chloroform. Chromatography was performed at 30 degrees C on a 40 X 3.2 mm I.D. C-18 guard cartridge (3 mu m spherical particles) using a mob ile phase of 11% (v/v) acetonitrile in 0.01 M phosphate buffer (pH 2.5 ) containing 0.001 M triethylamine, and pumped at 1 ml/min. Detection was at 279 nm. The retention times of norfloxacin and internal standar d were 1.9 and 2.9 min, respectively. Calibration curves were linear ( r > 0.999) from 0.1 mg/l to at least 2.0 mg/l. Within-day and between- day precision (C.V.) were 8.6% or less, and accuracy was 5.3% or less. Absolute assay recovery of norfloxacin was over 70%.