HIGH-RESOLUTION PENUMBRAL IMAGING OF 14-MEV NEUTRONS

Citation
Yw. Chen et al., HIGH-RESOLUTION PENUMBRAL IMAGING OF 14-MEV NEUTRONS, IEICE transactions on electronics, E78C(12), 1995, pp. 1787-1792
Citations number
12
Categorie Soggetti
Engineering, Eletrical & Electronic
ISSN journal
09168524
Volume
E78C
Issue
12
Year of publication
1995
Pages
1787 - 1792
Database
ISI
SICI code
0916-8524(1995)E78C:12<1787:HPIO1N>2.0.ZU;2-M
Abstract
We have developed a neutron imaging system based on the penumbral imag ing technique. The system consists of a penumbral aperture and a sensi tive neutron detector. The aperture was made from a thick (6 cm) tungs ten block with a toroidal taper. It can effectively block 14-MeV neutr ons and provide a satisfactory sharp, isoplanatic (space-invariant) po int spread function (PSF). A two-dimensional scintillator array, which is coupled with a gated two-stage image intensifier system and a CCD camera, was used as a sensitive neutron detector. It can record the ne utron image with high sensitivity and high signal-to-noise ratio. The reconstruction was performed with a Wiener filter. The spatial resolut ion of the reconstructed neutron image was estimated to be similar to 31 mu m by computer simulation. Experimental demonstration has been ac hieved by imaging 14-MeV deuterium-tritium neutrons emitted from a las er-imploded target.