FLOW-INJECTION AMPEROMETRIC DETERMINATION OF TRACE AMOUNTS OF AMMONIUM USING A GAS-DIFFUSION CELL AS THE SAMPLE LOOP

Citation
Rm. Liu et al., FLOW-INJECTION AMPEROMETRIC DETERMINATION OF TRACE AMOUNTS OF AMMONIUM USING A GAS-DIFFUSION CELL AS THE SAMPLE LOOP, Analyst, 120(12), 1995, pp. 2845-2848
Citations number
22
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032654
Volume
120
Issue
12
Year of publication
1995
Pages
2845 - 2848
Database
ISI
SICI code
0003-2654(1995)120:12<2845:FADOTA>2.0.ZU;2-6
Abstract
A new flow injection method for the determination of trace amounts of ammonium by using a pyrolytic graphite electrode electrodeposited with a cuprihexacyanoferrate film coated with Nafion film as a sensor is d escribed. A gas-diffusion unit was used to enhance the selectivity and sensitivity, Under optimum analytical conditions, the linear range of the calibration graphs for ammonium were 2-32, 0.8-16 and 0.4-8 mu mo l l(-1) when the accumulation time was 1, 2 and 3 min, respectively, T he method has been applied to the determination of ammonium in rain-wa ter samples and recoveries of 96.7-102.2% were obtained, The s(r) valu es of seven replicate analyses for all samples were 1.6-2.5%.