A DETECTOR FOR SUBMILLIMETER GAMMA-CAMERAS

Citation
Lm. Barone et al., A DETECTOR FOR SUBMILLIMETER GAMMA-CAMERAS, Nuclear physics. B, 1995, pp. 729-733
Citations number
6
Categorie Soggetti
Physics, Nuclear
Journal title
ISSN journal
05503213
Year of publication
1995
Supplement
44
Pages
729 - 733
Database
ISI
SICI code
0550-3213(1995):<729:ADFSG>2.0.ZU;2-H
Abstract
Anger cameras (SPECT etc.) presently used in nuclear medicine employ a s active detector NaI crystals, obtaining intrinsic spatial resolution s greater than or equal to 3mm. Arrays made of optically isolated sing le crystal elements of YAP:Ce, having sub-millimeter aperture size, re ad out by position sensitive photomultipliers, allow to build active d etectors to employ in SPECT systems, with intrinsic spatial resolution below the millimeter, and with time resolution of the order of tens o f nanoseconds. In this paper preliminary results of measurements carri ed on different kinds of YAP:Ce arrays are reported. The measurements have been performed aiming to optimize the geometrical and physical pa rameters of the crystals in order to accomplish a SPEM (Single Photon Emission Mammography) camera detector.