PSEUDO FRACTURE OF A DEFIBRILLATING LEAD

Citation
Jm. Kratz et al., PSEUDO FRACTURE OF A DEFIBRILLATING LEAD, PACE, 18(12), 1995, pp. 2225-2226
Citations number
3
Categorie Soggetti
Cardiac & Cardiovascular System","Engineering, Biomedical
ISSN journal
01478389
Volume
18
Issue
12
Year of publication
1995
Part
1
Pages
2225 - 2226
Database
ISI
SICI code
0147-8389(1995)18:12<2225:PFOADL>2.0.ZU;2-J
Abstract
Routine follow-up chest X ray of several patients with the Endotak(TM) transvenous pacing lead suggested fracture of the conductor at the di stal end of the proximal defibrillating coil. Electrical evaluation of these patients' leads demonstrated them to be working normally. Multi ple x-ray views of a new undamaged lead demonstrated that in some proj ections the Endotak lead may appear fractured in spite of normal conti nuity of the conductors. Physicians dealing with these patients should be aware of this confusing x-ray finding.