ACCURACY OF CERENKOV MEASUREMENTS USING A LIQUID SCINTILLATION SPECTROMETER

Citation
M. Takiue et al., ACCURACY OF CERENKOV MEASUREMENTS USING A LIQUID SCINTILLATION SPECTROMETER, Applied radiation and isotopes, 47(1), 1996, pp. 123-126
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
47
Issue
1
Year of publication
1996
Pages
123 - 126
Database
ISI
SICI code
0969-8043(1996)47:1<123:AOCMUA>2.0.ZU;2-V
Abstract
Cerenkov counting efficiency varies with colour quenching and sample t urbidity. The activity in a plastic vial can be determined accurately with a colour quenching correction technique, regardless of the presen ce or absence of turbidity in a sample. On the other hand, the error o f the measured activity in a glass vial becomes large with increasing sample turbidity due to the dissimilarity of the quench correction cur ves for non-turbid and turbid samples.