E. Grimvall et al., DETERMINATION OF POLYCHLORINATED-BIPHENYLS IN HUMAN BLOOD-PLASMA BY ONLINE AND OFF-LINE LIQUID-CHROMATOGRAPHY - GAS-CHROMATOGRAPHY, HRC. Journal of high resolution chromatography, 18(11), 1995, pp. 685-691
A method for the determination of non-, mono- and di-ortho substituted
polychlorinated biphenyls in human blood plasma is presented. The non
-ortho CBs are isolated from the bulk of PCBs by HPLC on a dinitroanil
ino-propyl silica column and collected as a heart-cut fraction that is
transferred on-line to GC-MS utilizing concurrent solvent evaporation
technique. The on-line coupling serves to lower the limit of determin
ation and to partly automate the sample clean-up. The di- and mono-ort
ho CBs are analyzed by off-line GC-ECD using two different columns. Sa
mple pre-treatment is performed by extraction with organic solvents an
d fractionation on silica gel. A high reproducibility and an absolute
over-all recovery of approximately 70% is demonstrated. The method is
applied to samples of blood from 5 individuals. All the samples demons
trated a similar relative concentration profile. It was also shown tha
t three CB congeners (CB 126, CB 156 and CB 170) contributed to the ma
jor part of the toxic equivalent quantity of the PCB content in the bl
ood.