A SIMULATION TO MODEL POSITION ENCODING MULTICRYSTAL PET DETECTORS

Citation
G. Tsang et al., A SIMULATION TO MODEL POSITION ENCODING MULTICRYSTAL PET DETECTORS, IEEE transactions on nuclear science, 42(6), 1995, pp. 2236-2243
Citations number
22
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
42
Issue
6
Year of publication
1995
Part
2
Pages
2236 - 2243
Database
ISI
SICI code
0018-9499(1995)42:6<2236:ASTMPE>2.0.ZU;2-E
Abstract
We have developed a simulation to model position encoding multicrystal detectors for positron emission tomography. The simulation is designe d to treat the interactions of energetic photons in a scintillator, th e geometry of the multicrystal array, as well as the propagation and d etection of individual scintillation photons. The simulation is tested with a model of the EXACT AR PLUS block detector manufactured by Siem ens-CTI, Position and energy responses derived from the simulation are compared to measured ones. Line spread functions, for four columns of crystals, are reproduced with an accuracy of +/- 0.5 mm, The crystal- by-crystal photopeak pulse heights and FWHM's are also predicted withi n a range of +/- 14%, and (-6)(+9)%, respectively.