REMOVAL OF IRON INTERFERENCE IN THE DETERMINATION OF SELENIUM IN GEOLOGICAL SAMPLES BY CONTINUOUS-FLOW HYDRIDE GENERATION ICP-AES

Citation
L. Martinez et al., REMOVAL OF IRON INTERFERENCE IN THE DETERMINATION OF SELENIUM IN GEOLOGICAL SAMPLES BY CONTINUOUS-FLOW HYDRIDE GENERATION ICP-AES, Fresenius' journal of analytical chemistry, 354(1), 1996, pp. 126-127
Citations number
17
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
354
Issue
1
Year of publication
1996
Pages
126 - 127
Database
ISI
SICI code
0937-0633(1996)354:1<126:ROIIIT>2.0.ZU;2-P
Abstract
To remove the chemical iron interference in the determination of selen ium in rocks, soils and stream sediments by Continuous Flow Hydride Ge neration Inductively Coupled Plasma Atomic Emission Spectrometry (CF-H G-ICP-AES), a Liquid-liquid extraction using ammonium pyrrolidine-1-yl -dithioformate (APDC) as a masking agent and chloroform as an extracti ng agent at pH 0.5-2.0 is proposed. Quantitative extraction of copper interfering seriously with the selenium determination by HG-ICP-AES is also achieved. Selenium could be accurately determined for selenium: iron ratios of 1 : 10(6), with a detection limit of 0.15 ng ml(-1). Ap plication of the method to Geochemical Standard Reference Samples demo nstrated that results were statistically in good accordance with certi fied values.