ABSOLUTE BRIGHTNESS MEASUREMENT OF X-RAY-EMISSION FROM LASER-PRODUCEDPLASMA USING THERMOLUMINESCENT DOSIMETERS

Citation
M. Farnikova et al., ABSOLUTE BRIGHTNESS MEASUREMENT OF X-RAY-EMISSION FROM LASER-PRODUCEDPLASMA USING THERMOLUMINESCENT DOSIMETERS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 368(2), 1996, pp. 484-487
Citations number
14
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
368
Issue
2
Year of publication
1996
Pages
484 - 487
Database
ISI
SICI code
0168-9002(1996)368:2<484:ABMOXF>2.0.ZU;2-Z
Abstract
Thermoluminescent dosemeters of LiF:Mg,Ti, LiF:Mg,Cu,P and CaF2:Dy wer e used as secondary standards in calibration measurements of X-ray emi ssion from a laser-produced aluminium plasma. Relative sensitivities o f the thermoluminescent dosemeters were found to be independent of dos e rate over a wide dose rate region. The photon detection efficiency o f thermoluminescent dosemeters is compared to that of semiconductor Si photodiodes. The thermoluminescent dosemeters are presented as a powe rful tool for testing of capability of a semiconductor photodiode dete ctor of X-ray pulse energy emitted from the laser-produced plasma.