BIOSENSORS BASED ON CROSS-LINKING OF BIOTINYLATED GLUCOSE-OXIDASE BY AVIDIN

Authors
Citation
Ms. Vreeke et P. Rocca, BIOSENSORS BASED ON CROSS-LINKING OF BIOTINYLATED GLUCOSE-OXIDASE BY AVIDIN, Electroanalysis, 8(1), 1996, pp. 55-60
Citations number
24
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
10400397
Volume
8
Issue
1
Year of publication
1996
Pages
55 - 60
Database
ISI
SICI code
1040-0397(1996)8:1<55:BBOCOB>2.0.ZU;2-F
Abstract
Avidin, having four biotin binding units, cross-links and immobilizes glucose oxidase (GOX) labeled with multiple biotins (B-GOX) when solut ions of B-GOX and avidin are mixed on an electrode. The H2O2 flux gene rated in the B-GOX catalyzed oxidation of glucose by dissolved O-2 is measured as an electroreduction current at a horseradish peroxidase (H RP) redox connducting hydrogel electrode poised at + 100 mV (Ag/AgCl). The sensitivity of the resulting glucose sensor is 0.14 A cm(-2) M(-1 ) with a linear range up to 2mM glucose.