OBSTRUCTION OF ARMORED TUBES USED IN PEDI ATRIC CIRCLE SYSTEMS

Citation
D. Hansen et al., OBSTRUCTION OF ARMORED TUBES USED IN PEDI ATRIC CIRCLE SYSTEMS, Anasthesist, 44(12), 1995, pp. 884-886
Citations number
9
Categorie Soggetti
Anesthesiology
Journal title
ISSN journal
00032417
Volume
44
Issue
12
Year of publication
1995
Pages
884 - 886
Database
ISI
SICI code
0003-2417(1995)44:12<884:OOATUI>2.0.ZU;2-I
Abstract
A separation of the individual latex layers is a known complication of endotracheal armoured tubes manufactured by immersion technique. This can result in herniation into the lumen with obstruction of the tube. Diffusion of nitrous oxide into the inner hernia considerably intensi fies the obstruction. This can lead to life-threatening situations. We observed a similar herniation caused by layer separation with subsequ ent tube obstruction of silicolatex anaesthesia tubes used in paediatr ic circuit systems. This is caused by manufacturing defects and above all by damage incurred in reprocessing. Therefore it is important to p rocess the tubes carefully. A too-high drying temperature can cause pr emature layer separation. The tubes should never be processed more tha n 80 times. Routine examination of the tubes is imperative, especially at the predilection sites for layer separation.