EFFECT OF AMMONIA ON THE RESPONSE OF THE ELECTRON-CAPTURE DETECTOR (ECD)

Citation
M. Abdelrehim et al., EFFECT OF AMMONIA ON THE RESPONSE OF THE ELECTRON-CAPTURE DETECTOR (ECD), HRC. Journal of high resolution chromatography, 18(12), 1995, pp. 757-759
Citations number
26
Categorie Soggetti
Chemistry Analytical
ISSN journal
09356304
Volume
18
Issue
12
Year of publication
1995
Pages
757 - 759
Database
ISI
SICI code
0935-6304(1995)18:12<757:EOAOTR>2.0.ZU;2-4
Abstract
The effect of ammonia on the electron capture detector (ECD) response has been investigated. Nitrogen with different ammonia concentrations (5-20%) was used as make-up gas. Compared to pure nitrogen, the ECD re sponse decreased when the ammonia concentration in the make-up gas was 5%, However, the response increased when the ammonia concentration wa s 20%. The response factor of 4-chlorophenol increased 4 fold when amm onia was 20%. Also, di- and tri-chlorophenols increased by 30-50%. The nitroaromatic compounds responses increased by about 23.7 times with 20% ammonia in the make-up gas. The signal-to-noise (S/N) increased wh en 20% ammonia in nitrogen was used as make-up gas compared to pure ni trogen. Also, the detector linearity increased by 50% with ammonia.