A PORTABLE FLOW-INJECTION ANALYZER FOR USE WITH ION-SELECTIVE ELECTRODES

Citation
T. Dimitrakopoulos et al., A PORTABLE FLOW-INJECTION ANALYZER FOR USE WITH ION-SELECTIVE ELECTRODES, Electroanalysis, 7(12), 1995, pp. 1118-1120
Citations number
15
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
10400397
Volume
7
Issue
12
Year of publication
1995
Pages
1118 - 1120
Database
ISI
SICI code
1040-0397(1995)7:12<1118:APFAFU>2.0.ZU;2-#
Abstract
A battery powered portable monitor based on flow injection potentiomet ry has been developed weighing 1.8 kg, constructed in a carry-case and applicable for remote site monitoring. Portability has been achieved using a light weight rechargeable Ni-Cd battery-pack (7.2 V). The flow injection manifold incorporates a low powered peristaltic pump, a wal l-jet type flow cell containing a commercial iodide ion-selective elec trode and a Ag/AgCl reference electrode connected to an analog-to-digi tal converter, also powered by the battery pack, and a RS232-serial ou tput to a notebook computer for real-time data display and storage. Th e system performance has been evaluated using the iodide ion-selective electrode, including the effects of various carrier solutions on samp le peak heights, peak widths and the calibration slopes and working ra nges. Fast response with peak widths as low as approximately 10 s. was observed with near Nernstian response of -53.8 mV change per activity decade and a log-linear range between 5 x 10(-6) and 1 x 10(-2) M in a 0.1 M sodium acetate carrier stream.