Electron energy-loss spectroscopy (EELS) provides a useful method for
determining the thickness of frozen-hydrated and dehydrated cryosectio
ns in terms of the inelastic mean free path. Cryosection thickness is
an important parameter because plural inelastic scattering limits the
sensitivity of elemental microanalysis based on core-loss EELS, and be
cause overlapping structures can affect interpretation of microanalyti
cal data as well as the quality of electron images. The purpose of thi
s work was to establish the minimum practical thickness for cutting cr
yosections and to explain the measured values for hydrated and dehydra
ted specimens. Hydrated sections were typically found to be between 1.
5-2.5 times thicker than expected from the nominal microtome setting;
this difference can be largely explained by compression during cutting
. Comparison of micrographs from hydrated and dehydrated cryosections
of rapidly-frozen, vitrified liver revealed a lateral shrinkage of sim
ilar to 20% on drying. The measured compression and shrinkage factors
are consistent with dark-field scanning transmission electron microsco
py (STEM) mass measurements on freeze-dried sections. Freeze-dried cry
osections, cut to a nominal thickness of 90 nm and supported on thin F
ormvar/carbon films, had a relative thickness t/lambda(i) in the range
of 0.5 for cytoplasm to 0.9 for mitochondria when analyzed at 100 keV
beam energy. Mass loss of similar to 30% occurring at high electron d
ose enabled useful core-loss spectra to be recorded even from high-mas
s compartments such as mitochondria without excessive plural scatterin
g. (C) 1996 Wiley-Liss, Inc.