MULTICENTER EVALUATION OF THE ABBOTT IMX(R) CA 15-3(TM) ASSAY

Citation
Gj. Vankamp et al., MULTICENTER EVALUATION OF THE ABBOTT IMX(R) CA 15-3(TM) ASSAY, Clinical chemistry, 42(1), 1996, pp. 28-33
Citations number
16
Categorie Soggetti
Chemistry Medicinal
Journal title
ISSN journal
00099147
Volume
42
Issue
1
Year of publication
1996
Pages
28 - 33
Database
ISI
SICI code
0009-9147(1996)42:1<28:MEOTAI>2.0.ZU;2-2
Abstract
We describe a new fully automated procedure for the quantitative measu rement of CA 15-3: the microparticle enzyme immunoassay (MEIA) technol ogy developed by Abbott Labs, for the IMx(R) automated immunoassay ana lyzer, The new IMx CA 15-3 test uses two mouse monoclonal antibodies, 115D8 and DF3. The test has a dynamic range to 250 kilounits/L and a m inimal detectable dose of CA 15-3 <0.2 kilounits/L. On dilution, linea rity is excellent, with recoveries ranging from 94% to 101%. Studies w ere conducted at four sites to evaluate the performance characteristic s of this assay, The intra- and interassay CVs were <4.7% and <5.6%, r espectively, and showed a between-laboratory CV <5.9%. Test results of the Abbott IMx CA 15-3 (y) were correlated with those obtained with t he Centocor(R) CA 15-3 RW (x), a solid-phase heterologous RIA. Linear regression analysis on results for 1973 samples yielded: y = 0.97x - 2 .09 (r = 0.9899, S-y\x = 22.2, range 1-4089 kilounits/L).