Jr. Mitchell et al., QUANTIFICATION OF MULTIPLE-SCLEROSIS LESION VOLUMES IN 1.5 AND 0.5 T ANISOTROPICALLY FILTERED AND UNFILTERED MR EXAMS, Medical physics, 23(1), 1996, pp. 115-126
Recently, guidelines for the use of MRI in the monitoring of MS have r
ecommended the use of imaging systems with mid-field (0.5-1.0 T) or hi
gh-held (greater than 1.0 T) strengths. Higher field strengths provide
many advantages, including increased signal-to-noise ratios (SNR). SN
R also may be increased by post-processing algorithms that reduce nois
e. In this paper we evaluate the impact on operator variability of (a)
lesion quantification in high-field (1.5 T) versus mid-field (0.5 T)
exams; and (b) an anisotropic diffusion filter algorithm that reduces
image noise without blurring or moving object boundaries. Inter- and i
ntra-operator reliability and variability were studied using repeated
quantification of lesions in 1.5 and 0.5 T filtered and unfiltered MR
exams of a MS patient. Results indicate that inter-operator variabilit
y in 1.5 T unfiltered exams was 0.34 cm(3) and was significantly large
r than that in 1.5 T filtered (0.27 cm(3)), 0.5 T unfiltered (0.26 cm(
3)), and 0.5 T filtered (0.24 cm(3)) exams. Similarly, intra-operator
variability in 1.5 T unfiltered exams was 0.23 cm(3) and was significa
ntly larger than that in 1.5 T filtered (0.19 cm(3)), 0.5 T unfiltered
(0.19; cm(3)), and 0.5 T filtered (0.18 cm(3)) exams. In addition, th
e minimum significant change between two successive measurements of le
sion volume by the same operator, was 0.64 cm(3) in 1.5 T unfiltered e
xams, but 0.53 cm(3) or less in other exams. For two different operato
rs making successive measurements, the minimum significant change was
0.94 cm(3) in 1.5 T unfiltered exams, but only 0.75 cm(3) or less in o
ther exams. Finally, the number of lesions to be monitored for an aver
age change in cm volume at a given power and significance level was gr
eater by 30%-60% for quantification in 1.5 T unfiltered exams. These r
esults suggest that inter- and intra-operator variability are reduced
by anisotropic filtering, and by quantification in 0.5 T exams. Reduce
d operator variabilities may result from higher detail signal-to-noise
ratios (dSNRs) in 0.5 T and filtered exams. (C) 1996 American Associa
tion of Physicists in Medicine.