The design of an analytical system based on the coupling of gas-diffus
ion separation and sequential injection analysis is described and appl
ied to ammonium determination. The sample and an alkaline solution are
sequentially aspirated using an automatic burette and mixed by flow r
eversal while they are being propelled to a gas-diffusion unit. There,
the ammonia formed diffuses through an hydrophobic membrane into an a
cid-base indicator solution. The change in the absorbance of the acid-
base indicator solution used as acceptor stream is measured and referr
ed to ammonium content in the sample. The system requiring simple and
commonly available instrumentation is automated and has been applied t
o the determination of ammonium in environmental samples.