Background. A totally implantable, intrathoracic electrohydraulic vent
ricular assist device (EVAD) is being developed for permanent use or a
s a bridge to transplantation. Methods. The blood pump with 70-mL nomi
nal stroke volume, volume displacement chamber, reversible turbine, in
ternal electronics and infrared diaphragm position sensor are combined
in one compact unit (unified system). The size and geometry are based
on human anatomic measurements and fluid dynamic studies. A transcuta
neous energy transfer powers the system and recharges the implantable
nickel-cadmium battery pack. Autotuning circuitry optimizes energy tra
nsfer efficiency over a range of transcutaneous energy transfer coil s
pacings and misalignments. An infrared diaphragm position sensor detec
ts end-systole and diastole points. Results. In vitro and acute in viv
o tests have demonstrated flow rates greater than 6 L/min. The transcu
taneous energy transfer system demonstrated power transfer efficiencie
s of 60% to 80% for power demands from 5 to 60 W. Thirteen systems are
currently undergoing durability testing; one has run for more than 75
0 days failure-free. The system recently sustained circulation in an a
cute calf implantation for 96 hours. Conclusions. Results of the in vi
tro and in vivo testing to date have demonstrated that the developed s
ystem can function effectively as a totally implantable ventricular as
sist device. Chronic in vivo evaluation is planned.