Peroxyacetyl nitrate (PAN) solutions in the concentration range 10-100
mu g PAN per mi solvent were distributed and analyzed by ten European
laboratories. Overall precision accounting for standards' preparation
, transport, storage and analysis resulted in standard deviations of 1
6 +/- 8% of the mean for the high- and 25 +/- 8% for the low-end of th
e concentration range. Following alkaline PAN hydrolysis, nitrite anio
ns were determined, mostly by chemically suppressed ion chromatography
, with better precision than the acetate anions.