T. Matsukawa et al., A COMPARISON OF 4 INFRARED TYMPANIC THERMOMETERS WITH TYMPANIC MEMBRANE TEMPERATURES MEASURED BY THERMOCOUPLES, Canadian journal of anaesthesia, 43(12), 1996, pp. 1224-1228
Purpose: To compare measurements made with four infrared tympanic ther
mometers (Genius(R), Thermopit(R), Quickthermo(R), and Thermoscan(R))
with those recorded from thermo couples positioned in the contralatera
l ear. Methods: Four tympanic thermometers were evaluated in 50 health
y volunteers (12 female and 38 male). Temperatures were measured, in r
andom order, at the right tympanic membrane four times and the highest
temperature was considered to be the true value measured by each ther
mometer. The control temperature was measured at the left tympanic mem
brane using Mon-a-Therm(R) thermocouples. Results: The tympanic membra
ne temperature measured by Genius(R) correlated best with the Mon-a-th
erm(R) measurement (TM) (r = 0.74). The tympanic membrane temperatures
measured by Thermopit(R), Quickthemo(R), and Thermoscan(R) correlated
moderately with TM (r = 0.56, 0.63, and 0.58, respectively). Mean dif
ferences between TM and each temperature (TG, TTP, TQ, and Trs) were -
0.3, 0.73, 0.42, and -0.3 degrees C, respectively. Likewise standard d
eviations were 0.33, 0.37, 0.35, and 0.35. Conclusion: We conclude tha
t all but the Thermopit(R) (TTP) are similarly useful for the manageme
nt of patients during anaesthesia.