W. Dusi et al., HGI2 X-RAY AND GAMMA-RAY CAMERA WITH SPECTROMETRIC CAPABILITIES, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 369(1), 1996, pp. 141-146
A HgI2 microstrip two-dimensional array with total area of 10 x 10 mm(
2) and a thickness of 0.5 mm has been tested. By using a coincidence-a
nticoincidence technique the effective pixel area has been identified
and spectroscopic performance and some indications about the spatial r
esolution have been extracted. Values like 25% (FWHM) at 6 keV and 8.5
% at 32 keV for the energy resolution have been obtained. For the X-ra
y energies used, ranging from 32 to 81 keV, the absence of an intense
lateral diffusion of the produced charge allows performing the positio
n readout in a quantized fashion, where the event position is that of
the struck strip, leading to a spatial resolution corresponding to the
pixel size.