Tw. Prendergast et al., SAFE APPLICATION OF DIAGNOSTIC LAPAROSCOPY DURING BIVENTRICULAR ASSISTANCE, The Annals of thoracic surgery, 61(2), 1996, pp. 735-737
We performed diagnostic laparoscopy in a patient who was critically un
stable while on an ABIOMED BVS 5000 biventricular assist system. A rel
atively unique feature of the ABIOMED is the venous return being depen
dent only on gravity and the drawing force of venous return being the
height of the inflow cannula compared with the level of the ABIOMED bl
ood pump assembly; this did not preclude the use of pneumoperitoneum n
ecessary for laparoscopy. The safe application of diagnostic laparosco
py in patients on a ventricular assist device is of importance in an e
ra of increasing use of these devices and the increased potential for
intraabdominal complications in this population.