THROMBOSIS OF MITRAL-VALVE BIOPROSTHESES IN PATIENTS REQUIRING CIRCULATORY ASSISTANCE

Citation
Mt. Hagley et al., THROMBOSIS OF MITRAL-VALVE BIOPROSTHESES IN PATIENTS REQUIRING CIRCULATORY ASSISTANCE, The Annals of thoracic surgery, 60(6), 1995, pp. 1814-1816
Citations number
8
Categorie Soggetti
Surgery,"Cardiac & Cardiovascular System
ISSN journal
00034975
Volume
60
Issue
6
Year of publication
1995
Pages
1814 - 1816
Database
ISI
SICI code
0003-4975(1995)60:6<1814:TOMBIP>2.0.ZU;2-5
Abstract
Two patients with new mitral valve bioprostheses required implantation of Bio-Medicus centrifugal pumps (Bio-Medicus, Minneapolis, MN) for c irculatory support and had evidence of prosthetic valve thrombosis 1 a nd 4 days later. Both patients died of thromboembolic complications de spite surgical removal of the thrombus. Thrombosis is a rare early com plication of bioprosthetic valves and in these cases was probably rela ted to low transvalvular now due to the use of circulatory assist devi ces. We discuss possible strategies for avoiding and managing this cat astrophic complication.