SIMPLE REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY QUANTITATION OF GANCICLOVIR IN HUMAN SERUM AND URINE

Citation
T. Page et al., SIMPLE REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY QUANTITATION OF GANCICLOVIR IN HUMAN SERUM AND URINE, Journal of chromatography B. Biomedical applications, 675(2), 1996, pp. 342-346
Citations number
8
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical applications
ISSN journal
15726495 → ACNP
Volume
675
Issue
2
Year of publication
1996
Pages
342 - 346
Database
ISI
SICI code
Abstract
A fast, simple, and cost-effective HPLC method for the quantitation of the antiviral drug ganciclovir is described. The serum samples are ex tracted with perchloric acid and neutralized with potassium phosphate buffer, and urine samples are diluted with distilled water. A reversed -phase column with isocratic elution by 15 mM potassium phosphate buff er (pH 2.5) containing 0.25% acetonitrile is used to separate ganciclo vir; quantitation is by UV absorbance at 254 nm. Total turnaround time is 22 min; more than 3000 samples can be run on a single column witho ut loss of peak quality. The limit of quantitation is 0.05 mu g/ml. Re coveries varied from 91 to 107% with coefficients of variation ranging from 0.387 to 7.95%.