M. Block et al., BIPHASIC DEFIBRILLATION USING A SINGLE CAPACITOR WITH LARGE CAPACITANCE - REDUCTION OF PEAK VOLTAGES AND ICD DEVICE SIZE, PACE, 19(2), 1996, pp. 207-214
The volume of current implantable cardioverter defibrillators (ICD) is
not convenient for pectoral implantation. One way to reduce the size
of the pulse generator is to find a more effective defibrillation puls
e waveform generated from smaller volume capacitors. In a prospective
randomized crossover study we compared the step-down defibrillation th
reshold (DFT) of a standard biphasic waveform (STD), delivered by two
250-mu F capacitors connected in series with an 80% tilt, to an experi
mental biphasic waveform delivered by a single 450-mu F capacitor with
a 60% tilt. The experimental waveform delivered the same energy with
a lower peak voltage and a longer duration (LVLD). Intraoperatively, i
n 25 patients receiving endocardial (n = 12) or endocardial-subcutaneo
us army (n = 13) defibrillation leads, the DFT was determined for both
waveforms. Energy requirements did not differ at DFT for the STD and
LVLD waveforms with the low impedance (32 +/- 4 Ohm) endocardial-subcu
taneous array defibrillation lead system (6.4 +/- 4.4 J and 5.9 +/- 4.
2 J respectively) or increased slightly (P = 0.06) with the higher imp
edance (42 +/- 4 Ohm) endocardial lead system (10.4 +/- 4.6 J and 12.7
+/- 5.7 J, respectively). However, the voltage needed at DFT was one-
third lower with the LVLD waveform than with the STD waveform for both
lead systems (256 +/- 85 V vs 154 +/- 51 V and 348 +/- 76 V vs 232 +/
- 54 V, respectively). Thus, a single capacitor with a large capacitan
ce can generate a defibrillation pulse with a substantial lower peak v
oltage requirement without significantly increasing the energy require
ments. The volume reduction in using a single capacitor can decrease I
CD device size.