F. Graichen et al., PATIENT MONITORING-SYSTEM FOR LOAD MEASUREMENT WITH SPINAL FIXATION DEVICES, Medical engineering & physics, 18(2), 1996, pp. 167-174
Fractures of the spine can be stabilized by different implants. Their
stiffness varies widely and only little is known about the loads actin
g on these devices. In order to measure the forces and moments in the
implant, the internal fixator after Dick was modified. An inductively
powered telemery unit is placed inside the fixator and is hermetically
sealed against body fluids. An integrated eight channel telemetry chi
p was developed to measure the signals of six strain gauge sensors, th
e implant temperature and the power supply. Because true fixators are
implanted together, two telemetry transmitters run at the sane time. T
his paper describes the function of the instrumented implant and the e
xternal system components.