Jl. Bernal et al., HPLC ANALYSIS OF CARBOHYDRATES IN WINES AND INSTANT COFFEES USING ANION-EXCHANGE CHROMATOGRAPHY COUPLED TO PULSED AMPEROMETRIC DETECTION, Journal of agricultural and food chemistry, 44(2), 1996, pp. 507-511
Carbohydrates (arabinose, fructose, fucose, galactose, glucose, mannos
e, rhamnose, ribose, sucrose, xylose, and the alditol mannitol) have b
een analyzed in wines and instant coffees, using anion exchange chroma
tography coupled to pulsed amperometric detection. Good separation and
resolution of the 11 compounds were obtained using just nanopure wate
r as eluent, although rinsing the column with 0.2 M NaOH was necessary
to avoid gradual decline in column resolution. Coffee solution, red w
ines, and rose wines needed a cleanup process by solid-phase extractio
n on C-18 cartridges. Arabinose, galactose, glucose, xylose, mannose,
fructose, and ribose were detected and quantified in red wines using m
elezitose as internal standard. In the case of the coffee samples mann
itol, fucose, arabinose, rhamnose, galactose, glucose, sucrose, mannos
e, and fructose were quantified and differences between unadulterated
and adulterated coffees could be detected.