As. Shanbhag et al., COMPOSITION AND MORPHOLOGY OF WEAR DEBRIS IN FAILED UNCEMENTED TOTAL HIP-REPLACEMENT, Journal of bone and joint surgery. British volume, 76B(1), 1994, pp. 60-67
Interfacial membranes collected at revision from 11 failed uncemented
Ti-alloy total hip replacements were examined. Particles in the membra
nes were characterised by electron microscopy, microchemical spectrosc
opy and particle size analysis. Most were polyethylene and had a mean
size of 0.53 mu m +/- 0.3. They were similar to the particles seen in
the base resin used in the manufacture of the acetabular implants. Rel
atively few titanium particles were seen. Fragments of bone, stainless
steel and silicate were found in small amounts. Most of the polyethyl
ene particles were too small to be seen by light microscopy. Electron
microscopy and spectroscopic techniques are required to provide an acc
urate description of this debris.