VOLTAGE DIVIDER FOR MEASURING MEGAVOLT PULSES WITH NANOSECOND RISE-TIME

Citation
An. Bastrikov et al., VOLTAGE DIVIDER FOR MEASURING MEGAVOLT PULSES WITH NANOSECOND RISE-TIME, Instruments and experimental techniques, 38(4), 1995, pp. 474-476
Citations number
4
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
00204412
Volume
38
Issue
4
Year of publication
1995
Part
1
Pages
474 - 476
Database
ISI
SICI code
0020-4412(1995)38:4<474:VDFMMP>2.0.ZU;2-P
Abstract
The divider consists of two electrically and magnetically coupled line s in a common shield. One is an oil-filled high-voltage line, the othe r is formed by an unshielded cable fixed on the wall inside the first line. The input of the second line is loaded to the identical shielded cable connected at the other end to a low-inductance shunt. The divis ion factor is determined by solving the electrostatic equation. The me asured factor (similar to 100) is in agreement with the calculation, a nd the divider faithfully reproduces the pulse shape. In a test experi ment, a 2-MV voltage pulse with a rise time of similar to 1 nsec was r ecorded.