A. Tinger et al., AN ANALYSIS OF INTRATREATMENT AND INTERTREATMENT DISPLACEMENTS IN PELVIC RADIOTHERAPY USING ELECTRONIC PORTAL IMAGING, International journal of radiation oncology, biology, physics, 34(3), 1996, pp. 683-690
Citations number
27
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: To evaluate the relative frequency and magnitude of intratrea
tment and intertreatment displacements in the patient positioning for
pelvic radiotherapy using electronic portal imaging. Methods and Mater
ials: Five hundred ninety-four electronic portal images of seven patie
nts treated with a four-field pelvic technique were evaluated. All pat
ients were treated prone without an immobilization device. Two fields
were treated per day, from which an average of two electronic portal i
mages were obtained for each field. No treatment was interrupted or ad
justed on the basis of these images. Each image was aligned to the cor
responding simulation film to measure the displacements in the mediola
teral, craniocaudal, and anteroposterior directions relative to the si
mulated center. The intertreatment displacement was the displacement m
easured from the initial image for each daily treated field. For each
daily treated field the intratreatment displacement was calculated by
subtracting the displacement measured on the initial image from the di
splacement measured on the final image. Results: The frequency of inte
rtreatment displacements exceeding 10 mm was 3%, 16%, and 23% for the
mediolateral, craniocaudal, and anteroposterior translations, respecti
vely. There were no intratreatment displacements exceeding 10 mm (p <
0.001). The frequency of intertreatment displacements exceeding 5 mm w
as 40, 52, and 51% for the mediolateral, craniocaudal, and anteroposte
rior translations, respectively; whereas, the frequency of intratreatm
ent displacements exceeding 5 mm was 1, 5, and 7% for the same transla
tions, respectively (p < 0.001). The standard deviation of the intertr
eatment displacements was at least three times as great as the standar
d deviation of the intratreatment displacements for all translations.
These deviations were greater than the precision limit of the measurem
ent technique, which is approximately 1 mm. Each patient had one direc
tion where systematic error predominated in intertreatment positioning
. Random error predominated for intratreatment positioning and for the
other two directions in intertreatment positioning. Conclusions: Duri
ng a course of pelvic radiotherapy, the frequency of intertreatment di
splacements exceeding 5 and 10 mm is significantly greater than the fr
equency of intratreatment displacements of these magnitudes. Errors in
intertreatment positioning are predominantly systematic in one direct
ion for each patient, whereas intratreatment error is predominantly ra
ndom. Because patients do not move considerably during the daily treat
ment of a pelvic field, a single electronic portal image per daily fie
ld may be considered representative of the treated position.