A HIGH-RESOLUTION AUTOGAIN RANGING LINEAR ELECTROMETER AMPLIFIER

Authors
Citation
Ss. Rajput et Sc. Garg, A HIGH-RESOLUTION AUTOGAIN RANGING LINEAR ELECTROMETER AMPLIFIER, Review of scientific instruments, 67(2), 1996, pp. 609-611
Citations number
3
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
67
Issue
2
Year of publication
1996
Pages
609 - 611
Database
ISI
SICI code
0034-6748(1996)67:2<609:AHARLE>2.0.ZU;2-Q
Abstract
A current electrometer, capable of operating over a wide range of temp eratures is presented. It measures bipolar currents from 3.16 pA to 50 mu A in ten sensitivity ranges on a linear scale with a current resol ution of 0.2%. The electrometer range depends upon the magnitude of th e input current. The measurement accuracy is better than +/-1% at 25 d egrees C for currents between 100 pA and 50 mu A and remains within +/ -3% in the temperature range of -15 to 55 degrees C. (C) 1996 American Institute of Physics.