AN IMPROVED LEAKAGE CURRENT COMPENSATION TECHNIQUE FOR A 4-PI-GAMMA IONIZATION-CHAMBER SYSTEM
Citation
Y. Hino et Kawada Y",nazaroh, AN IMPROVED LEAKAGE CURRENT COMPENSATION TECHNIQUE FOR A 4-PI-GAMMA IONIZATION-CHAMBER SYSTEM, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 369(2-3), 1996, pp. 392-396
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
SICI code
0168-9002(1996)369:2-3<392:AILCCT>2.0.ZU;2-T
Abstract
A current integration method using a small capacitor is most commonly
employed for precise measurements of small currents from 4 pi gamma io
nization chambers for the secondary standardization of radionuclides.
An improved technique has been developed for eliminating the possible
effect due to the electrical leakage and/or current loss across the fe
edback capacitor used in the integration of the ionization current. Th
is method is based upon charge integration from a fixed negative level
of potential to nearly the same level of positive potential via the z
ero point. The validity of this method is demonstrated for some typica
l applications of 4 pi gamma ionization chamber systems. This techniqu
e can contribute to the improvement of accuracy and to the extension o
f the intensity range of radioactive source in 4 pi gamma ionization m
easurement.