IMPROVEMENT OF THE CTA DOSIMETRIC PROPERTIES BY THE SELECTION OF READOUT WAVELENGTH AND THE CALCULATION OF THE SPECTROPHOTOMETRIC QUANTITY

Citation
F. Abdelrehim et al., IMPROVEMENT OF THE CTA DOSIMETRIC PROPERTIES BY THE SELECTION OF READOUT WAVELENGTH AND THE CALCULATION OF THE SPECTROPHOTOMETRIC QUANTITY, Applied radiation and isotopes, 47(2), 1996, pp. 247-258
Citations number
15
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
47
Issue
2
Year of publication
1996
Pages
247 - 258
Database
ISI
SICI code
0969-8043(1996)47:2<247:IOTCDP>2.0.ZU;2-X
Abstract
A detailed study was performed on sources of uncertainty associated wi th the measurement of cellulose triacetate (CTA) film dosimeter respon se at high absorbed hoses. The results show a significant dependence o f the uncertainty on the selected readout wavelength in the range 269- 290 nm, as well as the spectrophotometric quantity used for calculatin g the response. Two different calculated quantities were used, the mor e familiar one being the radiation;induced change in absorbance (Delta A)(i) = (A(i) - A(o))(i), of the film, for absorbance values before ( A(o)) and after irradiation (A(i)), and the other the ratio (A(i)/A(o) )(i) at a given wavelength. The effects of temperature, dose rate and r.h. of the surrounding atmosphere during and after irradiation on dos imeter performance, at some selected UV wavelengths of analysis e.g. l ambda = 274, 280 and 290 nm, were examined. Post-irradiation stability at different humidities was also studied. Improved environmental depe ndencies and dosimetric response properties were achieved by selecting the quantity(A(i)/A(o))(274nm).