EFFECT OF PRESSURE ON THE POTENTIOMETRIC RESPONSE OF ION-SELECTIVE ELECTRODES AND REFERENCE ELECTRODES

Citation
B. Muller et Pc. Hauser, EFFECT OF PRESSURE ON THE POTENTIOMETRIC RESPONSE OF ION-SELECTIVE ELECTRODES AND REFERENCE ELECTRODES, Analytica chimica acta, 320(1), 1996, pp. 69-75
Citations number
16
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
320
Issue
1
Year of publication
1996
Pages
69 - 75
Database
ISI
SICI code
0003-2670(1996)320:1<69:EOPOTP>2.0.ZU;2-6
Abstract
A study is presented on the influence of pressure of up to 30 bar on p otentiometric cells. Combinations of a poly(vinyl chloride) (PVC) matr ix nitrate ion-selective electrode with inner filling solution, a soli d contact PVC nitrate electrode, a capillary flow type reference elect rode, and a solid state reference electrode were tested. A linear incr ease of the potentials as a function of pressure was observed in all c ases after an initially non-linear phase. The overall potential change s in the investigated pressure range were between 0.2 mV and 1.5 mV, s lopes being typically 10 mu V bar(-1) and 40 mu V bar(-1) for nitrate electrodes with liquid filling solutions and for solid contact nitrate electrodes, respectively.