An integrated de superconducting quantum interference device (SQUID) m
agnetometer has been fabricated using a YBa2Cu3O7-x-SrTiO3-YBa2Cu3O7-x
multilayer process. The magnetometer consists of a 8.3 mmX8.6 mm pick
up coil coupled inductively to a small washer-type SQUID with 9 Ohm bi
crystal junctions via an integrated 10 1/2-turn input coil. Using simp
le direct-coupled read-out electronics with additional positive feedba
ck, a noise level of 9.7 fT/root Hz at 1 kHz and 53 fT/root Hz at 1 Hz
was measured inside a magnetically shielded room. Due to the low nois
e level biomagnetic heart and brain signals with high quality and the
very weak signals of the peripheral nerve system were recorded. (C) 19
96 American Institute of Physics.