INTEGRATED YBA2CU3O7-X MAGNETOMETER FOR BIOMAGNETIC MEASUREMENTS

Citation
D. Drung et al., INTEGRATED YBA2CU3O7-X MAGNETOMETER FOR BIOMAGNETIC MEASUREMENTS, Applied physics letters, 68(10), 1996, pp. 1421-1423
Citations number
22
Categorie Soggetti
Physics, Applied
Journal title
ISSN journal
00036951
Volume
68
Issue
10
Year of publication
1996
Pages
1421 - 1423
Database
ISI
SICI code
0003-6951(1996)68:10<1421:IYMFBM>2.0.ZU;2-L
Abstract
An integrated de superconducting quantum interference device (SQUID) m agnetometer has been fabricated using a YBa2Cu3O7-x-SrTiO3-YBa2Cu3O7-x multilayer process. The magnetometer consists of a 8.3 mmX8.6 mm pick up coil coupled inductively to a small washer-type SQUID with 9 Ohm bi crystal junctions via an integrated 10 1/2-turn input coil. Using simp le direct-coupled read-out electronics with additional positive feedba ck, a noise level of 9.7 fT/root Hz at 1 kHz and 53 fT/root Hz at 1 Hz was measured inside a magnetically shielded room. Due to the low nois e level biomagnetic heart and brain signals with high quality and the very weak signals of the peripheral nerve system were recorded. (C) 19 96 American Institute of Physics.