SELECTABLE-POWER PHOTOREACTOR FOR FLOW-INJECTION ANALYSIS SYSTEMS ANDHIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY POSTCOLUMN PHOTOCHEMICAL DERIVATIZATION

Citation
M. Lores et al., SELECTABLE-POWER PHOTOREACTOR FOR FLOW-INJECTION ANALYSIS SYSTEMS ANDHIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY POSTCOLUMN PHOTOCHEMICAL DERIVATIZATION, Journal of chromatography, 724(1-2), 1996, pp. 55-65
Citations number
17
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Volume
724
Issue
1-2
Year of publication
1996
Pages
55 - 65
Database
ISI
SICI code
Abstract
A new design for a post-column photochemical reactor is presented, dif fering from those described in the literature in that different workin g powers (between 8 and 40 W) may be selected in order to improve dete ction and linearity of response. The application of the device in post -column photoderivatization of phenolic aldehydes-both in high-perform ance liquid chromatography with UV-Vis or fluorescence detection (isoc ratic and gradient elutions) and in flow-injection analysis systems is discussed. Use of selectable-power photoreactors such as the one desc ribed here allows optimum conditions to be selected for each analyte. Moreover the coil length may be reduced.