A new method for fast data collecting in rotating-frame NQR imaging is
reported. A single unidimensional spin-density profile is composed by
this procedure, recording the evolution of the magnetization during t
he course of the nutation. Hence, a complete data set necessary for th
e reconstruction of a 1D profile is directly recorded by a single scan
provided by a train of short and intense radiofrequency pulses. A red
uction of one or even two orders of magnitude in the time necessary fo
r data acquisition is achieved. One-dimensional profiles and 2D spin-d
ensity images of geometrically simple objects are compared, using data
sets obtained by the present single-experiment and the previous multi
experiment methods. (C) 1996 Academic Press, Inc.