S. Bieri et al., REPRODUCIBILITY OF CONFORMAL RADIATION-THERAPY IN LOCALIZED CARCINOMAOF THE PROSTATE WITHOUT RIGID IMMOBILIZATION, Radiotherapy and oncology, 38(3), 1996, pp. 223-230
Citations number
31
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
The reproducibility of 3-dimensional (3D) conformal therapy in localiz
ed prostate cancer was studied in 14 patients, based on retrospective
analysis of 196 anteroposterior-posteroanterior (AP/PA) and lateral po
rtal images. The patients were treated supine without rigid immobiliza
tion using six isocentric coplanar conformal fields. Three different o
bservers independently compared the portal and simulation images, dete
rmining the deviation of each portal film from the corresponding simul
ation film. No significant deviations were observed in the cephalo-cau
dal or lateral axes (0 mm median values). However, a systematic median
shift of 5 mm (0 to +10, range) was observed in the anteroposterior d
irection, presumably as a consequence of a sagging in the treatment co
uch under the patient's weight. After modification of the treatment co
uch, no further systematic anteroposterior shifts have been observed.
These results demonstrate that the daily setup of conformal prostate i
rradiation fields can be performed with acceptable reproducibility wit
hout the use of special immobilization devices.