CALIBRATION OF SEMICONDUCTOR-DETECTORS FOR DOSE ASSESSMENT IN TOTAL-BODY IRRADIATION

Citation
N. Jornet et al., CALIBRATION OF SEMICONDUCTOR-DETECTORS FOR DOSE ASSESSMENT IN TOTAL-BODY IRRADIATION, Radiotherapy and oncology, 38(3), 1996, pp. 247-251
Citations number
9
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
01678140
Volume
38
Issue
3
Year of publication
1996
Pages
247 - 251
Database
ISI
SICI code
0167-8140(1996)38:3<247:COSFDA>2.0.ZU;2-Z
Abstract
The aim of this paper is to discuss the measurements carried out to im plement 'in vivo dosimetry' with EDP-30 diodes in total body irradiati on (TBI) techniques. Exit calibrations and calibrations behind cerrobe nd protection blocks showed the importance of calibrating diodes in al l relevant clinical conditions. Special attention was given to calibra tion of diodes behind cerrobend blocks. Dependence of the calibration factors on the thickness of the shielding blocks was, therefore, studi ed. This dependence was again studied after adding a wax cap to the di ode and when the ionisation chamber was placed at the same depth as th e measuring point of the diode. Temperature dependence in diode sensit ivity and dependence on accumulated dose for diodes response and for t emperature correction factors were also examined.