INSTRUMENTED IMPLANT FOR MEASURING TIBIOFEMORAL FORCES

Citation
Kr. Kaufman et al., INSTRUMENTED IMPLANT FOR MEASURING TIBIOFEMORAL FORCES, Journal of biomechanics, 29(5), 1996, pp. 667-671
Citations number
26
Categorie Soggetti
Engineering, Biomedical",Biophysics
Journal title
ISSN journal
00219290
Volume
29
Issue
5
Year of publication
1996
Pages
667 - 671
Database
ISI
SICI code
0021-9290(1996)29:5<667:IIFMTF>2.0.ZU;2-4
Abstract
Experimental measurement of loads occurring in the human knee joint wi ll allow validation of analytical models and provide data for the desi gn of total knee implants. A customized transducer was developed to me asure the dynamic tibiofemoral force and center of pressure after tota l knee arthroplasty. The transducer consists of a standard tibial comp onent to which four uniaxial load cells and an additional tibial tray have been added. The transducer was calibrated using a loading device traceable to the National Institute of Standards and Technology (NIST) . The transducer was accurate to within 1% in magnitude, 0.07 mm in me dial/lateral location and 0.24 m in anterior/posterior location.