QUANTITATIVE INJECTION FROM A MICROLOOP - REPRODUCIBLE VOLUMETRIC SAMPLE INTRODUCTION IN CAPILLARY ZONE ELECTROPHORESIS

Citation
Pk. Dasgupta et K. Surowiec, QUANTITATIVE INJECTION FROM A MICROLOOP - REPRODUCIBLE VOLUMETRIC SAMPLE INTRODUCTION IN CAPILLARY ZONE ELECTROPHORESIS, Analytical chemistry, 68(7), 1996, pp. 1164-1168
Citations number
16
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032700
Volume
68
Issue
7
Year of publication
1996
Pages
1164 - 1168
Database
ISI
SICI code
0003-2700(1996)68:7<1164:QIFAM->2.0.ZU;2-U
Abstract
A wire loop deployed at the tip of a capillary electrophoresis system has been investigated as a means of quantitative injection. A thin fil m of a liquid is formed on the loop, the loop is transferred to a seal ed chamber, and then pneumatic pressure is applied to introduce the co ntents of the loop into the capillary. As long as the applied pressure is below a certain threshold, no air is introduced into the capillary , even after the loop contents have been fully introduced. Sample surf ace tension and viscosity do not have a significant effect on the inje cted volume. The small loop injection technique appears to be a robust and reproducible alternative to presently practiced approaches to sam ple injection in CE.