Objective: To evaluate a hand held portable i-STAT analyzer used for p
oint-of-care testing in a hemodialysis unit. Design and Methods: Impre
cision and comparison studies. Assays by nonlaboratory personnel. Resu
lts: Study of both quality control solutions and hemodialysis patients
' duplicate blood specimens with 6+, E3+, EC6+, and EC8+ cartridges bo
th gave CVs for sodium, potassium, chloride, pH, and ionized calcium o
f 0 to 2.2%. CVs for hematocrit, glucose and pCO(2) for whole blood du
plicates were 4.6, 3.5, and 4.0%. The mean differences between i-STAT
and central laboratory values for the above compounds and urea, when e
xpressed as a % of mean central laboratory values were 0.2 and 3.1%, e
xcept for an ionized calcium value of 4.3% (for which there is an expl
anation) and for glucose (11%) that were both higher for i-STAT. Preli
minary data suggest that sodium, potassium, ionized calcium, and pH ca
n be assayed in the aqueous dialysate fluid used for hemodialysis. Con
clusions: We confirm previous reports that i-STAT cartridges give simp
le, rapid, and reliable values for sodium, potassium, chloride, urea,
glucose, and hematocrit that are clinically acceptable when performed
by nonlaboratorians, but also suggest the same conclusions apply to io
nized calcium, pH, and pco(2). The hemodialysis unit may prove to be o
ne of several hospital units where the analyzer has considerable value
.