Dj. Touw et al., EVALUATION OF A FLUORESCENCE POLAROGRAPHIC IMMUNOASSAY WITH INCREASEDSENSITIVITY FOR MEASUREMENT OF LOW CONCENTRATIONS OF TOBRAMYCIN IN SERUM, Therapeutic drug monitoring, 18(2), 1996, pp. 189-193
The limits of quantitation of the assay of tobramycin in serum by the
fluorescence polarization immunoassay system marketed by Abbott Labora
tories (TDxFLx system) are 0.1 and 10.0 mg/L. For some pharmacokinetic
studies, however, a more sensitive analysis is needed. The sensitivit
y of the TDxFLx system can theoretically be increased 10-fold by pipet
ting buffer solution into the sample well and 450 mu l serum into the
predilution well. The assay modified in this way can be run with the u
sual calibration of the apparatus for normal analysis. The analytical
performance of this modification of the TDxFLx assay was assessed. Tob
ramycin concentrations ranged from 0.01 to 1.0 mg/L. Analytical recove
ry ranged from 85 to 95%. The coefficients of variation for within-run
and between-run precision ranged from 0.5% to 5% and from 2% to 6%, r
espectively, for control concentrations ranging from 0.05 to 0.75 mg/L
. Based on recovery and precision results, the lower limit of quantita
tion was established as 0.025 mg/L. There was no significant detectabl
e cross-reactivity from ceftazidime, aztreonam, flucloxacillin, cilast
atin, or imipenem. There was a small, but significant cross-reactivity
from gentamicin and netilmicin. Hyperbilirubinemia did not affect the
assay, but hyperlipidemia gave falsely elevated results of the tobram
ycin assay. It was concluded that modification of the assay resulted i
n an acceptable method to quantify low concentrations of tobramycin in
serum.