VOLTAGE DIP IN PACEMAKER BATTERY SUPPLY - A NEW CAUSE OF PACEMAKER MEDIATED TACHYCARDIA

Citation
R. Stroobandt et al., VOLTAGE DIP IN PACEMAKER BATTERY SUPPLY - A NEW CAUSE OF PACEMAKER MEDIATED TACHYCARDIA, PACE, 16(4), 1993, pp. 806-811
Citations number
6
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
16
Issue
4
Year of publication
1993
Part
1
Pages
806 - 811
Database
ISI
SICI code
0147-8389(1993)16:4<806:VDIPBS>2.0.ZU;2-G
Abstract
A new cause of pacemaker mediated tachycardia was observed in a patien t equipped with a Vitatron(R) Quintech 931 DDD pacemaker. In this type of pacemaker the microprocessor is switched off during the atrial ref ractory period. Beyond the recommended replacement time, the internal resistance of the battery may increase to such an extent that switchin g on and off the microprocessor may cause voltage dips. Those voltage dips are erroneously interpreted as P waves by the atrial sensing ampl ifier, which may cause self-triggering of the pacemaker and initiate a pacemaker mediated tachycardia.