SHARING A PAIR OF ELECTRODES BETWEEN EXCITATION AND DETECTION IN A FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETER

Citation
G. Mauclaire et al., SHARING A PAIR OF ELECTRODES BETWEEN EXCITATION AND DETECTION IN A FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETER, Rapid communications in mass spectrometry, 10(5), 1996, pp. 591-593
Citations number
12
Categorie Soggetti
Spectroscopy,"Chemistry Analytical
ISSN journal
09514198
Volume
10
Issue
5
Year of publication
1996
Pages
591 - 593
Database
ISI
SICI code
0951-4198(1996)10:5<591:SAPOEB>2.0.ZU;2-H
Abstract
Time-sharing of electrodes between excitation and detection for Fourie r transform (FT) study of the coherent cyclotron motion of ions is dis cussed, together with the choice of a suitable commutation device. A p ractical implementation is described, and the application to an open-e nd, flat, ion cyclotron resonance cell presented. In addition, the swi tching technique described is useful to minimise ion losses due to res istive damping at the magnetron frequencies, and could also be used in conventional FT ion cyclotron resonance cells for quadrupolar excitat ion and detection, and for ion axialization.