ADVANTAGE OF MONITORING SKIN VIBRATION AS AN INDEX OF ARTERIOVENOUS FISTULAR BLOOD-FLOW
Citation
T. Aoki et al., ADVANTAGE OF MONITORING SKIN VIBRATION AS AN INDEX OF ARTERIOVENOUS FISTULAR BLOOD-FLOW, Artificial organs, 20(3), 1996, pp. 277-279
Categorie Soggetti
Engineering, Biomedical
SICI code
0160-564X(1996)20:3<277:AOMSVA>2.0.ZU;2-H
Abstract
Monitoring of the skin vibration is disturbed little by environmental
noise. Therefore, the skin vibration is more relevant than bruit on th
e skin by means of monitoring the fistular blood flow. Frequency analy
sis of the skin vibration generated by an internal arteriovenous fistu
la indicated two peaks of power spectral density (PSD), one in the fre
quency range of 4-10 Hz and the other from 100-300 Hz. The analysis of
the skin vibration generated by an external fistula, however, showed
only one peak in the range of 4-10 Hz and a moderate flat PSD level in
the range of 100-300 Hz. The PSD level in the range of 100-300 Hz dec
reased dramatically or disappeared when the fistular blood flow dimini
shed or ceased. Therefore, the optimal frequency for monitoring skin v
ibration may be the range of 100-300 Hz.