Abstract
There have been recent proposals of diffusion-weighting pulse sequence s that purport to measure the trace of the diffusion tensor, or to wei ght an image by that rotationally invariant quantity. Here the effect of a time-dependent diffusion tensor on such sequences is investigated theoretically. It is found that the time dependence of the diffusion tensor both breaks the rotational invariance of the pulse sequence and affects the overall magnitude of the diffusion weighting. These devia tions are calculated for two pulse sequences, with parameters relevant to experiments on brain tissue. The departure from rotational invaria nce is found to be surprisingly small: <1%, for reasonable parameters and similar to 5% for the most extreme choice of parameters. The overa ll magnitude of the diffusion weighting, however, may be significantly altered; 10% is estimated for reasonable parameters. It is proposed t hat this effect should prove a sensitive probe of time dependence in t he diffusion tensor. (C) 1996 Academic Press, Inc.